EasyManua.ls Logo

Megger MFT-X1 - R1 + R2 Mode

Megger MFT-X1
110 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
14.11 R1 + R2 mode
ATTENTION : You can only select R1+R2 on Hot key 1 when you have L-PE selected on Hot key 5.
This is because R1+R2 is only used for the phase to earth impedances of final circuit measurements.
You can only measure R1+R2 if you have measured Zref first (Zref is automatically stored when made).
1. Turn the left-hand rotary knob to the Loop Impedance test
position.
2. The display illuminates with a green background.
3. Plug your L (red) test-lead into the red socket and your PE
(green) test-lead into the green socket.
4. Ensure you have Hot key 5 in L-PE mode before repeatedly
pressing Hot key 1 to select R1+R2.
4.1. Alternatively, press Hot key 1 once and use the right-
hand rotary knob to select R1+R2.
REMEMBER: Hot key 5 MUST be in L-PE to allow R1+R2 to be
selected on Hot key 1 or R1+R2 will not appear as an option!
5. Repeatedly press Hot key 2 to select No RCD or RCD if one is
present.
5.1. Alternatively, press Hot key 2 once and use the right-
hand rotary knob.
6. Repeatedly press Hot key 4 to select either 3 wire, 2 wire or 2
wire HR testing. This example is using 2 wire.
6.1. Alternatively, press Hot key 4 once and use the right-
hand rotary knob.
7. Connect your test-leads to the L and PE conductors of the
supply at the point where it enters the building.
8. If using the 3 wire mode connect all three test-leads
9. Press and release the test button
If AUTO START is enabled in settings press the test button BEFORE
connecting the test-leads.
10. The Confidence Meter™ bar reduces towards the middle as
the circuit is analyzed.
11. When the test is complete, the display shows the R1+R2 value
and the Zref used to calculate it.
www.megger.com MFT-X1
73
Earth Loop Impedance testing – TrueLoop™

Table of Contents

Other manuals for Megger MFT-X1

Related product manuals