EasyManuals Logo

Mindray BeneVision TM80 Service Manual

Mindray BeneVision TM80
210 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #136 background imageLoading...
Page #136 background image
5-4 TM80 Telemetry Monitor Service Manual
5.4.1.6 ECG Calibration
The ECG signal may be inaccurate due to hardware or software problems. As a result,
the ECG wave amplitude becomes greater or smaller.
Tool required:
Vernier caliper
1. In the main menu, select Parameter Setup.
2. Select ECG.
3. Set Filter to Monitor.
4. Return to the main menu and select Maintenance.
5. Input the maintenance passcode.
6. Tap Accept.
7. In the Maintenance menu, select General.
8. Enable Calibrate ECG.A square wave appears on the screen and the message
ECG Calibrating is displayed in the technical alarm area of the devices screen.
Compare the amplitude of the square wave with the wave scale. The difference
should be within 5%.
9. After completing the verification, disable Calibrate ECG. If necessary, you can
print out the square wave and wave scale through the recorder and then
measure the difference.

Table of Contents

Other manuals for Mindray BeneVision TM80

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision TM80 and is the answer not in the manual?

Mindray BeneVision TM80 Specifications

General IconGeneral
BrandMindray
ModelBeneVision TM80
CategoryMonitor
LanguageEnglish

Related product manuals