TM80 Telemetry Monitor Service Manual 7-3 
 
7.1.4 The Working Duration of Battery Becomes Short 
Possible causes: 
  The lithium-ion battery is aged or is not fully charged. 
  AA batteries are not of the type specified by Mindray. 
  Wi-Fi signal is poor and frequent operation is performed on the screen.   
 
 
7.2 Technical Alarms 
 
Item  Alarm Message  Possible cause  Solution 
ECG  ECG Lead Off  The electrode has 
become detached from 
the patient or the lead 
wire has become 
disconnected from the 
adapter cable. 
Check the 
connections of the 
electrodes and 
leadwires. 
ECG XX** Lead 
Off 
ECG Module 
Error 
  An error occurred 
to the ECG module. 
  There is a problem 
with the 
communications 
between the 
module and the 
TM80. 
Restart the TM80. If 
the problem persists, 
contact your service 
personnel.