5-10 TM80 Telemetry Monitor Service Manual
You can either perform a manual leakage test:
1. Perform Steps 1 to-4 in the NIBP Accuracy Test section.
2. Raise the pressure in the rigid vessel to 250 mmHg with the balloon pump.
Then, wait for 5 seconds to let the measured values becoming stable.
3. Record the current pressure value and meanwhile use a time counter to count
time. Then, record the pressure value after counting to 60s.
4. Compare the two values and make sure the difference should not be greater
than 6 mmHg.
5.5 Miscellaneous Tests
5.5.1 Visual Inspection
Perform a visual inspection before the equipment is first used every day. Verify that
the equipment meets the following requirements:
The housing and display screen are free from cracks or other damages.
All keys function properly.
Connectors are not loose, cracked, or bent and cables have no cuts, nicks,
orfraying.
ECG leadwires are securely connected with the equipment.
Battery pack is installed and has sufficient charge.
Chest electrodes are free from cracks and limb electrodes can properly
clamp.
The external connectors are not loose and the pins are not bent.
The safety labels and data plates on the equipment are clearly legible.