TM80 Telemetry Monitor Service Manual 8-1
8 Disassembly
8.1 Overview
This chapter describes how to disassemble the TM80 and the BP10.
Before disassembling the TM80 and the BP10, make the following preparations:
Ensure that the TM80or the BP10is not used for monitoring a patient and
has been disconnected from the patient.
Disassemble the TM80 or the BP10 on the workbench of the equipment
maintenance room.
The personnel who will disassemble the TM80 or the BP10 must be
qualified for maintenance of medical equipment or have received
trainingsgiven by Mindray or have been authorized by Mindray.
In the disassembly process, the personnel disassembling the TM80or the
BP10 must take protective measures related toElectro-Static discharge
(ESD ).
NOTE
During re-installation after disassembling the device, the sealing strip
must be installed correctly. Improper installation of the sealing strip
may degrade the waterproof performance of the device.