EasyManuals Logo

Mindray BeneVision TM80 Service Manual

Mindray BeneVision TM80
210 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #206 background imageLoading...
Page #206 background image
A-14 TM80 Telemetry Monitor Service Manual
A.9 Scheduled Electrical Safety Inspection
For scheduled electrical safety inspection, perform all the test itemslisted
inA.11Electrical Safety Inspection Form.
A.10 Electrical Safety Inspection after Repair
The following table specifies test items to be performed after the equipment is
repaired. Refer toA.11Electrical Safety Inspection Formfor the description of the
test items.
For the TM80
Repair with main unit not disassembled
Test items: 1, 2,
Repair with main
unit disassembled
When patient
electrically-connected
PCBA is repaired or
replaced
Test items: 1.2.3.4.5
For the BP10
Repair with main unit not disassembled
Test items: 1, 2,
Repair with main
unit disassembled
When patient
electrically-connected
PCBA is repaired or
replaced
Test items: 1.2.

Table of Contents