A-4 TM80 Telemetry Monitor Service Manual 
 
A.4 Device Labeling 
Check the labels provided by the manufacturer or the healthcare facility are present 
and legible.   
  Main unit label 
  Integrated warning labels 
 
 
A.5 Earth Leakage Test 
Run an Earth Leakage test on the device being tested before performing any other 
leakage tests. 
 
Leakage current is measured the following ways:   
  Earth Leakage Current, leakage current measured through DUT outlet 
Earth   
  Earth Leakage Current AP-EARTH (ALL Applied Parts connected to Earth), 
leakage current measured through DUT outlet Earth   
There is no need to attach a test lead; the 601PRO automatically connects the 
measuring device internally.   
 
To Perform the Test 
1.  From the MAIN MENU, or with the outlet unpowered, plug the DUT into the 
601PRO front panel outlet, and turn on the device.   
2.  Attach the device's applied parts to the 601PRO applied part terminals if 
applicable. 
3.  Press shortcut key 4.The Earth Leakage test appears on the display, and the test 
begins immediately: