EasyManuals Logo

Mindray BeneVision TM80 Service Manual

Mindray BeneVision TM80
210 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #115 background imageLoading...
Page #115 background image
TM80 Telemetry Monitor Service Manual 3-91
3.11.2 Environmental Survey Table
After completing the environmental survey, fill in the survey results in the Check
Results column.
NO
Item Requirements of
TM80
Verification
Method
Check
Results
Wireless coverage requirements
1 Signal strength
(RSSI)
-65 dBm
Signal coverage
requirement for APs
which TM80 is
connected. RSSI value
is that perceived by
the TM80
Service person
performs the test
by using network
survey tool.
Make sure that all
expected coverage
areas such as ward,
corridor, toilet,
stairs, and elevator
are tested.
2 Co-channel
interference
-20dB
Measured on the
operating channel of
TM80
Service person
performs the test
by using network
survey tool.
Make sure that all
expected coverage
areas such as ward,
corridor, toilet,
stairs, and elevator
are tested.