EasyManua.ls Logo

Mindray BeneVision TM80 - Power Cord Plug

Mindray BeneVision TM80
210 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A-2 TM80 Telemetry Monitor Service Manual
A.2 Power Cord Plug
Test Item Acceptance Criteria
The power
plug
The power plug pins
No broken or bent pin. No discolored
pins.
The plug body No physical damage to the plug body.
The strain relief
No physical damage to the strain relief.
No plug warmth for device in use.
The power plug No loose connections.
The power cord
No physical damage to the cord. No
deterioration to the cord.
For devices with detachable power
cords, inspect the connection at the
device.
For devices with non-detachable power
cords, inspect the strain relief at the
device.

Table of Contents

Other manuals for Mindray BeneVision TM80

Related product manuals