A-2 TM80 Telemetry Monitor Service Manual 
 
A.2 Power Cord Plug 
Test Item  Acceptance Criteria 
The power 
plug   
The power plug pins 
No broken or bent pin. No discolored 
pins. 
The plug body  No physical damage to the plug body.   
The strain relief 
No physical damage to the strain relief. 
No plug warmth for device in use. 
The power plug  No loose connections. 
The power cord 
No physical damage to the cord. No 
deterioration to the cord.   
For devices with detachable power 
cords, inspect the connection at the 
device.   
For devices with non-detachable power 
cords, inspect the strain relief at the 
device.