TM80 Telemetry Monitor Service Manual A-3
A.3 Device Enclosure and Accessories
A.3.1 Visual Inspection
Test Item Acceptance Criteria
The enclosure and accessories
No physical damage to the enclosure
and accessories.
No physical damage to meters,
switches, connectors, etc.
No residue of fluid spillage (e.g., water,
coffee, chemicals, etc.).
No loose or missing parts (e.g., knobs,
dials, terminals, etc.).
A.3.2 Contextual Inspection
Test Item Acceptance Criteria
The enclosure and accessories
No unusual noises (e.g., a rattle inside
the case).
No unusual smells (e.g., burning or
smoky smells, particularly from
ventilation holes).
No taped notes that may suggest device
deficiencies or operator concerns.

Table of Contents