A-6 TM80 Telemetry Monitor Service Manual 
 
LIMITS 
For UL60601-1, 
  300 μA in Normal Condition 
  1000 μA in Single Fault Condition 
 
For IEC60601-1, 
  500 μA in Normal Condition 
  1000 μA in Single Fault Condition 
 
 
A.6 Patient Leakage Current 
Patient leakage currents are measured between a selected applied part and mains 
earth. All measurements have a true RMS only response.   
 
Preparation 
Perform a calibration from the Mains on Applied Part menu. 
 
The following outlet conditions apply when performing this test:   
  Normal Polarity, Earth Open, OutletON  Normal Polarity, OutletON 
  Normal Polarity, L2 Open, Outlet ON  Reversed Polarity, Outlet ON   
  Reversed Polarity, Earth Open, Outlet ON Reversed Polarity, L2 Open, 
OutletON