TM80 Telemetry Monitor Service Manual A-13
In Case of Failure
Check any broken of the enclosure. Replace any defective part.
Inspect wiring for bad crimps, poor connections, or damage.
Test the wall outlet; verify it is grounded and is free of other wiring
abnormalities. Notify the user or owner to correct any deviations. As a
work around, check the other outlets to see if they could be used instead.
Change another probe to confirm if the fail is caused by console.
If the leakage current measurement tests fail on a new unit and if situation
can not be corrected, submit a Safety Failure Report to document the
system problem. Remove unit from operation.
If all else fails, stop using and inform the Customer Service Engineer for
analysis and disposal.
LIMITS
For CF applied parts,
10μAin Normal Condition
50μAin Single Fault Condition
For BF applied parts,
100μA inNormal Condition
500μA in Single Fault Condition