7-4 TM80 Telemetry Monitor Service Manual
Item Alarm Message Possible cause Solution
ECG ECG Noise The ECG signal is noisy. Check for any
possible sources of
signal noise around
the cable and
electrode, and check
the patient for great
motion.
ECG Cable Type
Error
Connect wrong ECG
leadwire.
Reconnect the 3 or 5
lead ECG leadwire.
HR Overrange HR exceeds the
measurement limit.
Contact Mindray or
your service
personnel.
Resp Electrode Poor
Contact
The electrode has been
used for a long time or
the electrode contact is
poor.
Check the electrode
application.
Reposition
or replace the
electrodes if
necessary.
SpO
2
SpO
2
Sensor Off The SpO
2
sensor has
become detached from
the patient or the
module.
There is a fault with the
SpO
2
sensor.
Check the sensor
application site and
the sensor type, and
make sure the sensor
is not damaged.
Reconnect the sensor
SpO
2
Sensor
Fault
SpO
2
No Sensor

Table of Contents

Other manuals for Mindray BeneVision TM80

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision TM80 and is the answer not in the manual?

Mindray BeneVision TM80 Specifications

General IconGeneral
BrandMindray
ModelBeneVision TM80
CategoryMonitor
LanguageEnglish

Related product manuals