TM80 Telemetry Monitor Service Manual 8-7
10. Remove the speaker.
NOTE
The contact spring of the speaker is in the upper part.
11. Take down the ECG limit stop and remove the paddles.
NOTE
The gap of the ECG limit stop is on the right.
Speaker
Gap for ECG
limit stop