6 - 50 Operator’s Manual
6 Image Acquisition
The position for each plane is shown below:
6.15.4 Basic Procedures
Perform the following procedure:
1. Select a proper probe and a correct exam. Enter real-time B/Color/Power/Contrast mode (non-
cardiac Contrast Imaging mode).
3. Transverse
plane
R, L, A, and P imply the right, left, anterior and posterior of the
patient.
4. Coronal plane R, L, H, and F imply the right, left, head and foot of the patient.
5. Sagittal plane H, F, A, and P imply the head, foot, anterior and posterior of the
patient.
6.
Indicator
Four indicators (1
st
indicator is on positioning (probe) sensor and 2
nd
indicator is on motion (abdomen) sensor. 3
rd
indicator is on Needle
Navigation Guiding sensor. 4
th
indicator are reserved functions.)
Check the indicator becomes green in use.
Different indicator color:
• Blinks green: excellent.
• Blinks yellow: mediocre.
• Blinks red: poor.
• Stays grey: no signal activated. (Magnetic navigation is not
enabled.)
NOTE:
If the signal of the sensor becomes poor, check the connections of
the magnetic devices. Reset the distance/orientation among the
magnetic generator, sensors and the patient. Exclude the
interference from the metal devices (ward beds made of metal
materials) or the electromagnetic devices.
1 Coronal plane 2 Sagittal plane 3 Transverse plane
No. Image Description
1
2
3