12 Probes and Biopsy
Operator’s Manual 12 - 27
Perform the following procedure:
1. Install the needle-guided bracket:
a. Put on the sterile probe sheath.
b. Hold the probe by one hand, select the proper needle-guided bracket, and hold it with the
other hand. Align the narrow end tab of the needle-guided bracket with the groove of the
probe, then push the needle-guided bracket forward, making the tabs and the grooves of
the needle-guided bracket to match with the grooves and tabs of the probe.
c. Check manually to confirm that the needle-guided bracket is securely installed on the
probe.
2. Install the guiding block:
a. Select a proper guiding block and push it into the groove above the angle block, and
clamp it tightly.
b. Insert a biopsy needle with the same specification as that of the guiding block into the
hole of the guiding block.
3 Guiding block
4 Groove and tab of the needle-guided bracket
5 Specification of guiding block
6 Guiding hole of the biopsy needle
7 Specification of angle block