EasyManuals Logo

Mindray TE9 User Manual

Mindray TE9
312 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #201 background imageLoading...
Page #201 background image
12 Probes and Biopsy
Operator’s Manual 12 - 47
a. Insert a biopsy needle into the guiding block hole.
b. Adjust the biopsy angle of the needle, and ensure that the needle can be adjusted
continuously within the range of 30º to 85º.
3. Release the needle from the bracket:
Hold the probe, and remove the biopsy needle form the guiding hole of the guiding block.
4. Remove the needle-guided bracket:
a. Rotate the locking nut in the direction of the arrow shown on the figure to separate the
bracket from the probe.
b. Hold the probe and take out the bracket.
NGB-044 Metal-needle detachable needle-guided bracket
1Right clamp
2Groove
3 Guiding block fixing frame
4 Locking nut of left and right clamps
5 Guiding blocks
6 Depth shift sign
7 Depth pinch nut
8 Left clamp
1234
5678

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TE9 and is the answer not in the manual?

Mindray TE9 Specifications

General IconGeneral
BrandMindray
ModelTE9
CategorySpeaker System
LanguageEnglish