22
Calibration
2
1
3
2
3
After setting, push key to display .
After setting, push key to display .
F
ENTER
* The weighing capacity memorized now is displayed.
Select the changed digit.
Increment the value of selected digit.
The weighing
capacity
memorized now.
Set-up weighing capacity
Weight value equal to
weighing capacity
Weight value put on
measurement section
Step
5
Step
6
Set the weighing capacity
* The mass of the weight memorized now is displayed.
Set the weighing capacity.
Set the mass of the weight.
Set the mass of the weight
Select the changed digit.
CAL-ZCAL-Z
Increment the value of selected digit.
CAL-SCAL-S
Set the mass of weight
Calibration of ZERO
▲
▲
When key is pushed at display,
is displayed.
is displayed.
When key is pushed at display,
F
ENTER
1
F
ENTER
F
ENTER
CAL-ZCAL-Z
CAL-SCAL-S
display blinks for about 2 seconds when the calibration is executed, and [Weighing
capacity < Mass of weight] is set.
Please set the each value to become as [Weighing capacity ≧ Mass of weight]
Error display for setting the weighing capacity or mass of the weight.
• Please adjust the mass of the weight used for the calibration to 2/3 or more of the
weighing capacity.
• When the weight of the same mass as weighing capacity can be prepared,
weighing capacity is equal to mass of weight.
• When the SPAN calibration in [step 8] is registered by the value input, set as
[Weighing capacity equal to mass of weight]