33
Calibration
4-10-3. ZERO mode
ZERO calibration is executed by reading the load cell output value at ZERO by pushing
CAL-ZCAL-Z
key for 2 seconds or more.
1
By pushing key for 2 seconds or more from the normal
measurement mode,
blinks.
2
* Please make the condition that nothing is put on the
measuring section.
After
blinks, push the key.
3
ZERO is memorized after
blinks for 2 seconds,
and you can return to the measurement mode.
Measurement mode
Measurement mode
STABLE
ENTERENTER
FF
CAL-ZCAL-Z
Push for 2
seconds or more
CAL-ZCAL-Z
blinks for about 2 seconds when the output and input value of load cell
exceeds over ZERO adjustment range at minus side (at -2.5 mV/V or less)
Blinks for about 2 seconds when the output and input value of load cell
exceeds over ZERO adjustment range at plus side (at 2.5 mV/V or more.)
* Please conrm the rated capacity and the tare amount, etc., so that ZERO should not exceed
the adjustable range.
keeps the display when the value of [Registered SPAN + Registering ZERO
again] is over 3.1 mV/V.
* Please register SPAN continuously or repeat the registration of ZERO again.
Error display of ZERO calibration