USE MANUAL
AVIOR-2/AVIOR-2 I/O & MIP-2/MIP-2 I/O
106491_C User manual AVIOR-2 MIP-2_En.docx
MIRION TECHNOLOGIES (Canberra) S.A.S, BP249, ZI de Vauzelles, 3760
Counting scale informations
Measure state R integer16 unique IR 80
Integrated measure value R Float32 unique
IR 81 : MSB
IR 82 : LSB
Unit : available in IR 85
Integration time R Float32 unique
IR 83 : MSB
IR 84 : LSB
Measure unit R integer16 unique IR 85
1 : BQEQ
2 : BQEQ_PER_CM2
3 : SV_EQ_PER_H
4 : SV_PER_H
5 : CPS (US)
6 : CPM
7 : DPM
8 : DPM_PER_100CM2
9 : RPH
10 : REMPH
11 : UNKNOWN
Uncertainty R Float32 unique
IR 86 : MSB
IR 87 : LSB
Integrated counts R Float32 unique
IR 88 : MSB
IR 89 : LSB
Counting scale alarm
level
R integer16 unique IR 90
1 : LOW
2 : HIGH
3 : SATURATION
4 : COUNT
Only available at the end of
the measure
Channel 2
Measure state R integer16 unique IR 100
0 : NOT_STARTED
1 : IN_PROGRESS
Integrated measure value R Float32 unique
IR 101 : MSB
IR 102 : LSB
150,2
Unit : available in IR 105
Integration time R Float32 unique
IR 103 : MSB
IR 104 : LSB
30
Unit : seconds