23 STORAGE AND TRANSPORT
23.1 STORAGE
The AVIOR-2 or MIP-2 should be kept as far as possible from shock, crushing, and any other element
that may affect its integrity. It must also be kept out of the weather and under environmental conditions
corresponding to its storage area (- 25°C to + 60°C).
During storage, remove the battery from the AVIOR-2 and MIP-2. Refer to § 2.
: During storage, the battery of the AVIOR-2 and MIP-2 requires recharging every 18 months.
23.2 TRANSPORT
The packaging of the AVIOR-2 and MIP-2 for transportation must be sized to protect it from shock and
shock and to be compatible with standard delivery conditions.
24 PREVENTIVE MAINTENANCE
24.1 PERIODIC TABLE
The following table shows the preventive maintenance operations to perform:
Operation Periodicity Procedure
Verification that the AVIOR-2 is working properly
18 months
in storage
See § 23.2.1
1 year
After each
opening of
the AVIOR-
2
Verification that the MIP-2 is working properly
18 months
in storage
See § 23.2.2
1 year
After each
opening of
the MIP-2
Battery charging
18 months
in storage
See § 9
According
to need in
use