24.2.2 MIP-2
Functionality to check Procedure
Detection of the CSP probes by
the device.
- Connect a CSP probe. The display should read « Search »
followed by « Probe found »
Communication with CSP probe - Connect a CSP probe. Wait for initialization and check that the
device displays the measurement
Detection of the Nardeux probes
by the device.
- Connect a Nardeux probe. The display should read « Search »
followed by « Probe found »
Communication with Nardeux
probe
- Connect a Nardeux probe. Wait for initialization and check that
the device displays the measurement
Powered by the sector - Check that the charging indicator lights up during the power
connection.
Powered by USB - Check that the charge indicator lights up during USB connection.
Key detection on MIP-2 - Open the menu, and go to « USB mode », then « Connect a
USB key ». (See § 14.2)
- Check the communication with the USB key.
MIP-2 detection on PC - Open the menu, and go to « USB mode », then « Connect to a
computer » (See § 14.1).
- Check that the AVIOR-2 data is accessible from the PC..
Loud speaker sound - Make a long press on the left button to open the sound setting
(see § 13.4).
- Vary the volume to check the operation of the speaker.
Navigation buttons - Browse the menu using at least once the 5 buttons to check
their operation.
Backlight - Make a long press on the right button to check the correct
operation of the backlight (see § 13.3).
MIP-2 I/O
24V power supply - Check that the charge indicator lights up when connecting 24V.
Digital I/O check - Check the state of the relays with an ohmmeter (see § 19.1).
Receipt of data by the RS485
- Connect the AVIOR-2 to a PC via the RS485 (See §19.2.).
- Open a software using the ModBus protocol and then
interrogate the device.
Headphones sound - Connect the headphones to the jack..
- Make a long press on the left button to open the sound setting.
- Vary the volume to check the operation of the headphones.
25 DECONTAMINATION AND CLEANING
AVIOR-2 and MIP-2 are easily decontaminable.
The AVIOR-2 and MIP-2 have an IP54 protection rating which protects them from dust deposits and
splashing and splashing water.
Do not use stripper, wax, or corrosive solvents to clean the front of the unit, use only a soft, dry or damp
cloth.