106491_C User manual AVIOR-2 MIP-2_En.docx
MIRION TECHNOLOGIES (Canberra) S.A.S, BP249, ZI de Vauzelles, 3760
Press the key several times to return to the main menu.7.
14.11 INTERMEDIARY PERIODICAL CONTROL (IPC) OF AVIOR-2 OR MIP-2
The configuration of the IPC is only accessible at the « Administrator » user level (To change the user
level, refer to § 13.15).
Press the key to access the main menu1.
Note: The « Measure Mode » line is selected by default. If necessary press the or keys to
select the « Settings » line
Press the or key to view the « Settings » screen.2.
Note: The line « Background noise » is selected by default. Press the or keys to
select the « Periodic Control » line.
Press the or key to display the « Periodic control » menu.3.
Note: The « Probes » line is selected by default. Press the key or to select the
« Monitor » line.