USER MANUAL
AVIOR-2/AVIOR-2 I/O & MIP-2/MIP-2 I/O
106491_C User manual AVIOR-2 MIP-2_En.docx
MIRION TECHNOLOGIES (Canberra)S.A.S, BP249, ZI de Vauzelles, 37602
Loches Cedex
51 / 118
13 CONFIGURATION OF PROBES
13.1 CONFIGURING LOW / HIGH ALARM THRESHOLDS AND UNIT
The threshold values of low alarm, the threshold alarm high and the units of the connected probe are
accessible from the user « Technician » level.
User level « Operator » has only the ability to change units.
For probes of the CSP type, 10 threshold values by unit of measure are pre-recorded in the probe. The
thresholds can be changed without possibility of adding.
The following table shows an example for a CSP probe (SAB 100):
Channel Į Low or high alarm thresholds
4 Bq 8 Bq 40 Bq 80 Bq 400 Bq 800 Bq 2 kBq 8 kBq 40 kBq 80 kBq
400 mc/s 800 mc/s 4 c/s 8 c/s 40 c/s 80 c/s 200 c/s 800 c/s 4 kc/s 8 kc/s
0.04
Bq/cm
2
0.08
mBq/cm
2
0.4
mBq/cm
2
0.8
mBq/cm
2
4 Bq/cm
2
8 Bq/cm
2
20
Bq/cm
2
80
Bq/cm
2
400
Bq/cm
2
800
Bq/cm
2
Channel ȕ Low or high alarm thresholds
40 Bq 80 Bq 200 Bq 400 Bq 800 Bq 2 kBq 4 kBq 8 kBq 20 kBq 40 kBq
1 c/s 16 c/s 40 c/s 80 c/s 160 c/s 400 c/s 800 c/s 1.6 kc/s 4 kc/s 8 kc/s
0.4 Bq/cm
2
0.8 Bq/cm
2
2 Bq/cm
2
4 Bq/cm
2
8 Bq/cm
2
20
Bq/cm
2
80
Bq/cm
2
200
Bq/cm
2
400
Bq/cm
2
140
kBq/cm
2
For Nardeux type probes, the 10 threshold values for all units are recorded in the MIP-2. (See § 12.2).
The thresholds can be modified from the menu, without the possibility of adding any.
They are accessible from the main menu (key ), by selecting the line « Configuration of probes »
(key ).
In the case of 2 connected probes, press the or key to select the measurement 1.
interface.