12 USER MESSAGES
Messages on screen Action
« Access denied »
Unauthorized function for current user level, change
user level (§ 13.16)
« Excessive background noise »
In « Counting scale » mode § 10.4 and « Hand /
Foot » § 10.6, the measurement time is greater than
or equal to the maximum integration time.
Check the background noise parameters (§13.1)
In « Direct measurement » mode § 10.3,
« Screening » § 10.5, and « Hand / foot »
(according to configuration) § 10.6, the decision
threshold SD is greater than or equal to the alarm
threshold set in the probe.
Check the background noise parameters (§13.1)
Check the alarm thresholds of the probe (§12.1)
« BGK expired
Validate to recalculate
Return to exit »
Perform a background noise measurement.
Check the background noise settings (§ 13.1).
« IPC expired »
The probe is no longer usable and it is imperative to
carry out an Intermediate Periodic Inspection
12.3).
« Interface 2
Intermediate Periodic Inspection
expired »
The probe remains operational but it is advisable to
carry out an intermediate periodic check (§ 12.3).
« CPC expired »
The probe is no longer usable and it is imperative to
perform a periodic calibration check with the
Canberra Smart Probe Software (CSPS) software.
« Interface 2
Periodic calibration check
expired »
The probe remains operational but it is advisable to
perform a periodic calibration check with the
Canberra Smart Probe Software (CSPS) software.)
« Monitor Error
IPC expired »
The AVIOR-2 / MIP-2 is no longer usable and it is
imperative to carry out an intermediate periodic
inspection 13.11).
« Monitor: Intermediate Periodic Inspection
expired »
The AVIOR-2 / MIP-2 remains operational but it is
advisable to carry out an intermediate periodic
inspection (§ 13.11)
« Monitor error
CPC expired »
The AVIOR-2 / MIP-2 is no longer usable and it is
imperative to perform a periodic calibration check
13.12)
« Monitor: Periodic calibration check
expired »
The AVIOR-2 / MIP-2 remains operational but it is
advisable to perform a periodic calibration check
13.12)