EasyManua.ls Logo

Mirion Technologies AVIOR-2 - Acquisition of Background Noise in Net Mode

Mirion Technologies AVIOR-2
120 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
USER MANUAL
AVIOR-2/AVIOR-2 I/O & MIP-2/MIP-2 I/O
106491_C User manual AVIOR-2 MIP-2_En.docx
MIRION TECHNOLOGIES (Canberra)S.A.S, BP249, ZI de Vauzelles, 37602
Loches Cedex
43 / 118
Press the key or to display the « Measure Mode » and press the key to select 2.
the line « Measure Type: …. ».
The end-of-line character string indicates « Net » for a measure with deduced BGK or « Gross »
for a measure without deduced BGK.
Press or to display « Net » or « Gross ».
11.7.1 Acquisition of background noise in « Net » mode
When the « Measure Type: Net » is selected. An acquisition of background noise is then necessary.
Two types of acquisition can be performed:
- Dynamic acquisition when the probes are equipped with presence sensors. The presence sensor
determines if there is a potentially contaminated item to be measured. Otherwise the unit will able
to background noise during the minimum integration time setting (see § 13.1).
The device is then ready to perform a « NET » type measurement and waits for a presence:

Table of Contents