In case of two probes connected, press or key to select "Interface 1" or "Interface 2" 4.
line.
Note: For MIP-2, only interface 2 is compatible.
Press the or key to display the choice of the periodic control5.
Note: The « Intermediate Control » line is selected by default. Press the or keys to
select the line « Calibration Control »
Press the or key to display the « Calibration control » menu.6.
Note: the activation of the measurement lock activates automatically the notification
To activate the notification of the expired control and / or block the measurement:
x Select the line with or ,
x Activate the box with the key ,
x Exit with the key and / or select another line to be modified with the or keys.