USER MANUAL
AVIOR-2/AVIOR-2 I/O & MIP-2/MIP-2 I/O
106491_C User manual AVIOR-2 MIP-2_En.docx
MIRION TECHNOLOGIES (Canberra)S.A.S, BP249, ZI de Vauzelles, 37602
Loches Cedex
57 / 118
Record activity and uncertainty.8.
For example the SAB100 probe has:
x a beta efficiency (over 2ʌRIZLWKDQXQFHUWDLQW\RIN IRUWKH
60
Co,
x an alpha efficiency (over 2ʌRIZLWKDQXQFHUWDLQW\RINIRU
239
Pu.
With the following formula for yield: ܴ=
்௫ି஻ௗி
ʣ
× 100
With:
x Tx: Count rate (c / s)
x BoF: Count rate of background noise (c / s)
x Ɏ: Fluency rate of the source updated and corrected by a surface correction coefficient
Press the key to save the measurement and save the validity date or press the key to 9.
restart the measurement in step 6.
To change the duration:
x Increase or decrease the number with or ,
x Change column with or ,
x Save the duration with the key .
Select another channel (Step 4) or press the key to return to the « Intermediate Check »10.
screen (Step 3) and check the validity date of the IPC.
Press the key several times to return to the « Main Menu » screen11.