EasyManua.ls Logo

MR tapcon 240 - Visual display; Setting delay time T1; Setting control response T1

MR tapcon 240
174 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7 Functions and settings
Maschinenfabrik Reinhausen 201470 222/08 ENTAPCON® 240
Visual display
The deviation from the set bandwidth is shown visually in the device's dis-
play. The measured voltage highlighting shows whether the measured
voltage is above, within or below the set bandwidth . Progress of delay
time T1 is indicated by the gradual filling of the time bar
. The seconds
display above this indicates the remaining delay time T1.
Figure 43: Visual display of deviation from desired value
1 Bandwidth (upper and lower
limit)
4 Desired voltage value U
Ref
2 Time bar for delay time T1 5 Remaining delay time T1
3 Measured voltage U
Act
Setting delay time T1
Use this parameter to set delay time T1. This function delays the issuing of a
tap-change command for a defined period. This prevents unnecessary tap-
change operations if the tolerance bandwidth is exited.
To set the delay time T1, proceed as follows:
1. > Parameter > Control parameter > Press until the
desired parameter is displayed.
2. Press to highlight the position.
ð The desired position is highlighted and the value can be changed.
3. Press to increase the time or to reduce it.
4. Press .
ð The delay time T1 is set.
Setting control response T1
The control response T1 can be set to linear or integral.
With linear control response, the device responds with a constant delay time
regardless of the control deviation.
7.4.2.3
7.4.3
7.4.4
Linear control response T1

Table of Contents

Other manuals for MR tapcon 240

Related product manuals