Error and warning messages
403.xxx Warning [403.xxx] Check CT SB2 switch box wiring: CTA OUTPUT, TAPS XN and/or X1 may be
disconnected.
Reason: The CTA OUTPUT sockets of the CT SB2 are not properly connected to the CT Analyzer
output.
Solution: Make sure that the CTA OUTPUT sockets on the CT SB2 are properly connected to the
CT Analyzer and restart your test.
404.xxx Warning [404.xxx] Check CT SB2 switch box wiring: CTA PRIM and/or PRIM IN terminal may be
disconnected.
Reason: The CTA PRIM sockets of the CT SB2 are not properly connected to CT Analyzer input
PRIM.
Solution: Make sure that the CTA PRIM sockets on the CT SB2 are properly connected to
CT Analyzer input PRIM and restart your test.
405.xxx Warning [405.xxx] Check CT SB2 switch box wiring: CTA SEC terminal, TAPS XN IN and/or X1 IN
may be disconnected.
Reason: The CTA SEC sockets of the CT SB2 are not properly connected to CT Analyzer input
SEC.
Solution: Make sure that the CTA SEC sockets on the CT SB2 are properly connected to
CT Analyzer input SEC and restart your test.
406.xxx Warning [406.xxx] Check CT SB2 switch box wiring: <text depending on connection actually
detected>
Reason: The CTA SEC sockets of the CT SB2 are not properly connected to CT Analyzer input
SEC.
Solution: Make sure that the CTA SEC sockets on the CT SB2 are properly connected to
CT Analyzer input SEC and restart your test.
407.xxx Warning [407.xxx] Check CT SB2 switch box wiring: Wrong polarity at CTA SEC or CTA OUTPUT
terminal
Reason: Reverse polarity at the CTA SEC sockets or the CTA OUTPUT sockets of the CT SB2.
Solution: Make sure that the CTA SEC sockets and the CTA OUTPUT sockets on the CT SB2 are
properly connected to CT Analyzer and restart your test.
408.xxx Warning [408.xxx] Check CT SB2 switch box wiring: <text depending on missing connection
actually detected>
Reason: At least one of the CT taps enabled for multi-ratio testing is not connected to the CT SB2
correctly.
Solution: Make sure that all CT taps are connected correctly to the TAPS terminals on the CT SB2
and restart your test.
409.xxx Warning [409.xxx] Check CT SB2 switch box wiring: <text depending on incorrect connection
actually detected>
Reason: Wiring of at least two CT taps is mixed, e.g. CT tap X2 is connected to TAPS X3 on the
CT SB2 and vice versa.
Solution: Make sure that all CT taps are connected to the correct TAPS terminals on the CT SB2
and restart your test.
411.xxx Warning [411.xxx] Firmware download to switch box failed! Error: SBErrNo
Reason: CT SB2 firmware download failed.
Solution: Ensure that the CT SB2 is connected to the CT Analyzer and repeat the firmware
down
load to the CT SB2.
415.xxx Warning [415.xxx] Quick measurement cannot be started. Please disconnect the CT SB2 switch
box.
Reason: You are trying to initiate a Quick measurement with the CT SB2 switch box still connected
to the CT Analyzer.
Solution: Measurements using Quick Test are not possible when the CT SB2 is connected to the
CT Analyzer. Remove all connections to the CT SB2 and restart your Quick measurement
on the CT Analyzer.

Table of Contents

Other manuals for Omicron CT SB2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omicron CT SB2 and is the answer not in the manual?

Omicron CT SB2 Specifications

General IconGeneral
BrandOmicron
ModelCT SB2
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals