CT Analyzer User Manual
180 OMICRON
501.xxx No message displayed on the CT Analyzer. Message only displayed on remote PC connected to the
CT Analyzer.
Reason: The remote interface handler couldn't decipher the command.
Solution: Remote interface error. Check the connection between CT Analyzer and PC.
504.xxx No message displayed on the CT Analyzer. Message only displayed on remote PC connected to the
CT Analyzer.
Reason: Data transmission error.
Solution: Check the connection between CT Analyzer and PC.
505.xxx No message displayed on the CT Analyzer. Message only displayed on remote PC connected to the
CT Analyzer.
Reason: Data transmission error.
Solution: Check the connection between CT Analyzer and PC.
510.xxx No message displayed on the CT Analyzer. Message only displayed on remote PC connected to the
CT Analyzer.
Reason: Remote interface error: Failed to get the required measurement data because the
measurement is still running.
Solution: Check the connection between CT Analyzer and PC.
511.xxx No message displayed on the CT Analyzer. Message only displayed on remote PC connected to the
CT Analyzer.
Reason: Remote interface error: Failed to unzip the transmitted data block.
Solution: Check the connection between CT Analyzer and PC.
513.xxx No message displayed on the CT Analyzer. Message only displayed on remote PC connected to the
CT Analyzer.
Reason: The current remote interface command is not allowed.
Solution: Make sure that the current CT Analyzer state allows the execution of the command.
514.xxx No message displayed on the CT Analyzer. Message only displayed on remote PC connected to the
CT Analyzer.
Reason: Remote interface error: Remote command was submitted with an invalid parameter.
Solution: Verify the parameter (index) submitted with the remote interface command.
800.xxx Warning [800.xxx] Flash card access error. Internal file system error.
Reason: The file system sent an error message.
Solution: Compact Flash card is possibly corrupt. Use a new Compact Flash card.
801.xxx Warning [801.xxx] Flash card access error. Invalid file name or path.
Reason: The file system sent an error message.
Solution: The Compact Flash card is possibly corrupt. Try to save the data to a PC and format the
CF card.
802.xxx Warning [802.xxx] Flash card access error. Access denied.
Reason: The file system sent an error message. You tried to open a "read only" file or a special
directory.
Solution: Access the Compact Flash card on a PC and remove the "read-only" file attribute. File
attributes cannot be modified with the CT Analyzer.
803.xxx Warning [803.xxx] Flash card access error. File or folder already exists.
Reason: The file system sent an error message.
Solution: Enter a different file or folder name.
804.xxx Warning [804.xxx] Flash card access error. Disk full.
Reason:
The file system sent an error message.
Solution: Delete some files on the Compact Flash card or insert a new CF card.
805.xxx Warning [805.xxx] Flash card access error. No or corrupt flash card.
Reason: No or no valid CF card can be found.
Solution: Insert a valid Compact Flash card.

Table of Contents

Other manuals for Omicron CT SB2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omicron CT SB2 and is the answer not in the manual?

Omicron CT SB2 Specifications

General IconGeneral
BrandOmicron
ModelCT SB2
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals