10. MAINTENANCE
In the event an error occurs, an error message appears at the bottom of LCD display. Make sure of its
nature. Sections 10.1 Error Messages and 10.2 Status Messages will assist you in identifying the cause
of trouble and finding necessary corrective action. If the problem persists, or if you have any question,
contact the factory.
10.1 Error Messages
Error
type
Status
(Status display)
Description Coping action
I
ndications in error
Influences on output in an error
RED
LED
GRN
LED
Backlight
(※1
Analog
Output 1
Analog
Output 2
Pulse
Output 1
Pulse
Output 2
Status
Output
Saturated
Alarm
Analog Output 1
Saturated
(Ana. Out 1 Satur)
A
nalog output 1 out of
1
0 to 110(%) range; failure
to produce proper output.
Measurement is out of range. Adjust measurement
such that measurement selected falls within [lower
limit of analog output to upper limit value], or modify
setting. (6)
Blink ON OFF
21.6mA None
Unaffected
Per
status
select
(7)
2.4mA None
Analog Output 2 Saturated
(Ana. Out 2 Satur
Analog output 2 out of
10 to 110(%) range; failure
to produce proper output.
N
one
2
1.6mA
None 2.4mA
Pulse Output 1 Saturated
(Pls. Out 1 Satur)
Pulse output 1 exceeds 11kHz; failure to produce
proper output.
Instant flowrate selected exceeds full scale flowrate
of pulse output. Lower instant flowrate for use or
increase full scale flowrate setting. (5)
Unaffected
11KHz
None
Pulse Output 2 Saturated
(Pls. Out 2 Satur)
Pulse output 2 exceeds 11kHz; failure to produce
proper output.
None
11KHz
Sensor
Failure
Drive Input Out of Range
(Drive OutOfRange)
Drive frequency out of normal range; failure to make
proper measurement.
A fault in drive oil connections is a possibility. Verify
drive coil connections by self diagnostics.
(For details,see 9.6.2 Drive coil check.)
ON OFF
ORG
(
4)
Per error output
Setting
(6)
P
er error
output
S
etting
(5)
P
er
status
s
elect
(7)
Scale Over
(Scale Over)
Mass flowrate exceeds 110% of max. allowable
range; possible failure to make proper measurement.
Mass flowrate exceeds the specification.
Use within the maximum allowable range.
Temperature Out of
Range
(Temp OutOfRange)
Temperature out of normal range, failure to make
proper measurement.
A fault in temperature sensor is a possibility. Verify
temperature sensor connections by self diagnostics.
(For details,see 9.6.3 Transmitter check.)
Density Outside Limits
(Density Outside)
Density measured out of normal range (0 to 5[g/mL],
failure to make proper measurement.
Bubble entrapment is a possibility.
Check for bubble entrapment.
P.O. Sig Error(P.O. Sig
Err)
Pickoff signal voltage out of normal range, failure to
make proper measurement.
A
fault in pickoff sensor is a possibility. Verify pickoff
coil connections by self diagnostics.
(For details,see 9.6.3 Transmitter check.)
T
emperature Connect Error
(Temp Connect Err)
A
fault possible in temperature sensor wiring
connections
A
fault in temperature sensor is a possibility. Check
temperature sensor connections.
P
.O. Connect Error
(P.O. Connect Err)
A
fault possible in pickoff sensor wiring connections
A
fault in pickoff coil connections.
Check pickoff coil connections for condition.
Drive Coil Error A fault possible in drive coil wiring connections
A fault in drive coil connections. Check drive coil
c
onnections. (This error does not show up in
measurement. Upon error diction, be sure to verify
d
rive coil connections by self diagnostics.)
Transmitter
Failure
EEPROM Error
(EEPROM Err)
An error in parameters; inactive operation A fault in EEPROM is suspected. Contact the factory.
Blink
(2)
OFF
ORG
(4)
2.4mA Stopped OFF
Data Update Error
(Data Update Err)
A
fau
l
t in internal data
A fault in transmitter is suspected. Contact the
factory.
O
N
O
FF
ORG
(4)
Per error output
setting (3, 6)
Per error output
setting (3, 5)
Status
Select (7)
P
arameter
Alarm
Analog 1 Set Alarm
(Ana. 1 Set Alm.)
Parameter setting of analog output 1 or analog
output 2 comes under any of the following conditions
Upper limit < lower limit
Upper limit or lower limit > max. acceptable setting
Upper limit or lower limit < max. acceptable setting
V
erify parameters and reconfigure to acceptable
values. (6)
B
link
O
N
O
FF
Per error output
s
etting
(6)
Per error
output
s
etting
(5)
S
tatus
Select
(7)
Analog 2 Set Alarm
(Ana. 2 Set Alm.)
H
/L Alarm Point Set
Alarm
(H/L po.Set Alm.)
H/L alarm parameter setting comes under any of the
f
ollowing conditions
High alarm < Low alarm
High alarm or low alarm > max. acceptable setting
High alarm or low alarm < max. acceptable setting
Verify parameters and reconfigure to acceptable
v
alues. (For details, see 9.16 H/L Alarm Functions.)
Calibration
F
ailure
Auto Zero Failed
(Auto Zero Fail)
Auto zero offset (correction) is out of normal range.
Wait until the fluid flow settles down to a steady state
and retry zeroing.
Blink OFF OFF Unaffected Unaffected
Status
Select (7)
Slug Flow
A
larm
Slug Flow Alarm
(Slug Flow)
S
tays on for a preset duration (slug flow check) upon
detention of gas entrapment
Air entrapment in the fluid is a possibility. Verify the
fluid and line for conditions.
Blink ON OFF
O
nly flow output
Clamped
Clamped
S
tatus
Select
(7)
A
fter expiration of slug flow check duration upon gas
entrapment detection
F
low output at
"
0" S
topped
Transmitter
Alarm
X
mtr Temperature Alarm
(Txr Temp Alm.)
T
ransmitter
s
internal temperature is unusual (above
80°).
High ambient temperature cause a rise in the
t
ransmitter
s
internal temperature, leading to shorten
life of components. Take necessary step to lower
a
mbient temperature.
Blink ON OFF Continued Continued
Status
S
elect
(7)
Switch Alarm(Switch Alm.)
Faulty operating switches Faulty switch operation. Contact the factory. OFF ON OFF Continued Continued
1: Even with the run indicator staying off, the white backlight comes on in response to switch operation when backlight is set in the "ON" or "sleep".
2: In an alarm, the LED blinks at a faster rate (at intervals of 100ms) than in other alarms (at intervals of 250ms).
3: Outputs can go uncontrolled under certain circumstances if the transmitter fails, however.
4: For 5 minutes after occurrence of an alarm condition (or sleep duration of backlight setting), the ORG backlight comes on; after which it will blink on and off.
5: For details,see 9.12 Pulse Output Function.
6: For details,see 9.13 Analog Output Function.
7: For details,see 9.14 Status Output Function.