3.2.2 Type S general performance
Stainless steel tube type (CS010 to CS080)
Item Description
Model CS010 CS015 CS025 CS040 CS050 CS080
Flowrate
Guar. min. meas. rate (kg/h)
36 120 360 900 1800 3600
Min. setting rate (kg/h) 72 240 720 1800 3600 7200
Max. service rate (
kg/h
)
720 2400 7200 18000 36000 72000
Max. allowable rate (kg/h)
1080 3600 10800 27000 54000 108000
Accuracy ±0.2% ± Zero stability error of reading
Repeatability ±0.1% ± 1/2 Zero stability error of reading
Zero stability (kg/h) 0.36 1.2 3.6 9 18 36
Analog output accuracy 0.1% of full scale for respective accuracy
3.2.3 Type B general performance
Item Description
Model CB006 CB010 CB015 CB025 CB040 CB050
Flowrate
Guar. min. meas. rate (kg/h)
24 76.8 192 576 1920
Min. setting rate (kg/h) 60 192 480 1440 4800
Max. service rate (kg/h) 600 1920 4800 14400 48000
Max. allowable rate (kg/h) 1200 3840 9600 28800 96000
Accuracy ±0.2%±Zero stability error of reading
Repeatability ±0.1%±1/2 Zero stability error of reading
Zero stability (kg/h) 0.09 0.288 0.72 2.16 7.2
Density
Measuring rate 0.3 to 2.0g/mL
Accuracy (Option) ±0.003 g/mL
Analog output accuracy 0.1% of full scale for respective accuracy
Titanium tube type (CS010 to CSR50)
Item Description
Model CS010 CS015 CS025 CS040 CS050 CSR50
Flowrate
Guar. min. meas. rate (kg/h)
36 120 360 900 1800
Min. setting rate (kg/h) 72 240 720 1800 3600
Max. service rate (
kg/h
)
1220 4080 12200 30600 61200
Max. allowable rate (kg/h)
1830 6120 18300 45900 91800
Accuracy ±0.15% ± Zero stability error of reading
Repeatability ±0.075% ± 1/2 Zero stability error of reading
Zero stability (kg/h) 0.12 0.41 1.22 3.06 6.12
Density
(liquid)
Measuring rate 0.5 to 1.0g/mL, 0.7 to 1.3g/mL, 1.0 to 1.5g/mL (※1
Accuracy (Option) ±0.002g/mL
Analog output accuracy 0.1% of full scale for respective accuracy
1Density range varies with the fluid used.
※:When measuring the density at fluid temperature 40 or higher, heat retention measures must be taken.
※:The general performance is based on factory calibration accuracy.
NOTES
NOTE: The general performance is based on factory calibration accuracy.
NOTE: The general performance is based on factory calibration accuracy.