L
--
740
--
28
--
E
1
High temperature service models (CA025 to CA080)
Item Description
Model
CA025
(High temp. service)
CA040
(High temp. service)
CA050
(High temp. service)
CA080
(High temp. service)
Flowrate
(liquid)
Guar. min. meas. rate
(kg/h)
108 390 1200
Min. setting rate (kg/h) 540 1950 6000
Max. allowable rate
(kg/h)
21600 78000 240000
Max. service rate (
kg/h
)
10800 39000 120000
Accuracy
[±0.1% ± Zero stability error] of reading
Repeatability
[±0.05% ± 1/2 Zero stability error] of reading
Zero stability (kg/h) 1.08 3.9 12
Density
(liquid)
Measuring rate 0.3 to 2.0g/mL
Factory calib. acc. (opt) ±0.003g/mL
Analog output accuracy 0.1% of full scale for respective accuracy
High temperature service models (CA100, CA150)
Item Description
Model
CA100 (High temp. service) CA150 (High temp. service)
Flowrate
(liquid)
Guar. min. meas. rate
(kg/h)
3420
Min. setting rate (kg/h) 17100
Max. service rate (
kg/h
)
342000
Max. allowable rate
(kg/h)
684000
Accuracy
[±0.1% ± Zero stability error] of reading
Repeatability
[±0.05% ± 1/2 Zero stability error] of reading
Zero stability (kg/h) 34.2
Density
(liquid)
Measuring rate 0.3 to 2.0g/mL
Factory calib. acc. (opt) ±0.003g/mL
Analog output accuracy
0.1% of full scale for respective accuracy
NOTE: The general performance is based on factory calibration accuracy.
NOTE: The general performance is based on factory calibration accuracy.