EasyManua.ls Logo

Philips BiPAP AVAPS - Performing Testing Using the Service Center Manual Test Software; 1. Before starting the test, perform the following (refer to Figure 8-10):; Figure 8-10: Manual Calibration Initial Setup; 2. Turn on the manometer and verify that it is functioning properly.

Philips BiPAP AVAPS
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PAGE 8-10
1118824, VER. 00
PERFORMING TESTING USING THE SERVICE CENTER MANUAL TEST SOFTWARE
The Final Manual Test must be performed after repairs have been made to the device, the RTC has been
calibrated, and the device has been run in for two hours.
1. Before starting the test, perform the following (refer to Figure 8-10):
Connect the device to com port 1 (COM1) of your computer via the Link Module and DB9F-
DB9M Cable (if not already connected). Refer to Section 1.3 for information on connecting the
device to a PC.
Connect the O
2
Enrichment Attachment (occluded) to the Outlet Port of the Device.
Attach the Pressure Tubing to the O
2
Enrichment Attachment and the manometer.
FIGURE 8-10: MANUAL CALIBRATION INITIAL SETUP
2. Turn on the manometer and verify that it is functioning properly.
3. Launch the Service Center Manual Test from the Windows Start menu. Refer to Figure 8-11.
NOTE
You must clear the device’s error log before and after testing is performed.
Device Connected to Computer via Link
Module and DBPF-DB9M Cable

Table of Contents

Other manuals for Philips BiPAP AVAPS

Related product manuals