3. Turn the Blower “On”.
4. Turn the UI Knob to any setting other than “0” and verify the following:
The Humidifier Icon is displayed, and
The Heater Plate gets warm.
5. Record the result on the test data sheet on page 15.
8.5 READING/VERIFYING THE DEVICES SERIAL NUMBER AND MODEL NUMBER
To read or verify the serial number and model number of the device, perform the following:
1. Connect the device to a PC. Refer to Chapter 1 for information on connecting the device to a PC.
2. Apply power to the device.
3. Open the Service Center Tools Suite from the Windows Start Menu.
4. Select Read Serial Number & Model Number from the drop-down menu.
5. Click on the Execute Tool button.
6. The serial number and model number of the device will be displayed.
NOTE
The Read Serial Number & Model Number tool can be used only on the NIV
devices, not the System One Heated Humidifier.