6.1.15 REPLACING THE ALARM ASSEMBLY AND/OR ALARM HOLDER
FIGURE 6-20: ALARM ASSEMBLY AND ALARM HOLDER
To remove the Alarm Assembly and Alarm Holder:
1. Remove the UI Knob. Refer to Replacing the User Interface (UI) Knob.
2. Remove the Top Cover. Refer to Replacing the Top Cover.
3. Remove the Main PCA. Refer to Replacing the Main PCA.
4. Remove the Blower Cap. Refer to Replacing the Blower Cap.
5. Remove the Blower Assembly. Refer to Replacing the Blower Assembly and/or the Blower Outlet
Bellows.
6. Remove the Right Panel Assembly. Refer to Replacing the Right Panel Assembly.
7. Remove the Humidifier Cable. Refer to Replacing the Humidifier Cable.
8. Remove the Blower Housing. Refer to Replacing the Blower Housing.
9. Lift the Alarm Assembly and Housing out of the Bottom Enclosure.
10. Remove the Alarm Assembly from the Alarm Holder.
To install the Alarm Assembly and Alarm Holder:
1. Place the Alarm Assembly into the Alarm Holder as shown in Figure 6-20.
2. Route the Alarm Assembly Wires through the loop in the Alarm Holder as shown in Figure 6-20.
3. Place the Alarm Assembly and Alarm Holder into the Bottom Enclosure as shown in Figure 6-20.
Included in Kit Tools Required
Alarm Assembly T8 Torx screwdriver
T15 Torx screwdriver
Included in Kit Tools Required
Alarm Holder T8 Torx screwdriver
T15 Torx screwdriver