Philips BiPAP AVAPS - 7.2.4 Replacing the Dry Box Assembly;Humidifier Inlet Seal; Figure 7-6: Blower Cap; 1. Remove the Humidifier Tank Assembly. Refer to.; 2. Insert a flat blade screwdriver into the hole located on the back of the Humidifier and lightly press inward to release the Dry Box Assembly. Refer to Figure 7-7.

Philips BiPAP AVAPS
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PAGE 7-71118824, VER. 00
7.2.4 REPLACING THE DRY BOX ASSEMBLY/HUMIDIFIER INLET SEAL
FIGURE 7-6: BLOWER CAP
TO REMOVE THE DRY BOX ASSEMBLY/HUMIDIFIER INLET SEAL:
1. Remove the Humidifier Tank Assembly. Refer to .
2. Insert a flat blade screwdriver into the hole located on the back of the Humidifier and lightly press
inward to release the Dry Box Assembly. Refer to Figure 7-7.
3. Remove the Dry Box Assembly with Inlet Seal.
Included in Kit Tools Required
Dry Box Assembly
Inlet Seal
Dry Box Seal
Flathead Screwdriver
CAUTION
Do not press firmly on the screwdriver as damage to
the Humidifier may occur.
Dry Box Assembly
Inlet Seal