PAGE 4-8
1118824, VER. 00
FIGURE 4-8: BIPAP S/T PROVIDER MENU
Mode CPAP, S, S/T, T, or PC. The default setting is S/T. PC mode is only available on the BiPAP
AVAPS device.
Flex Control ““no” Allows the user to adjust the Bi-Flex setting if in “S” mode. “Yes” locks the Bi-Flex
setting so users cannot adjust it by selecting “yes”.
Bi-Flex
(FLEX)
The setting of Off, 1, 2, or 3 is the level of air pressure relief that the patient feels when
exhaling during therapy. The default setting is Off.
CPAP CPAP pressure setting from 4-20 in 1.0 increments. The default setting is 10.
AVAPS Enable or disable AVAPS when in S, S/T, T, or PC mode by selecting “yes” or “no”. The
default setting is “no”. This setting is available only on the BiPAP AVAPS device.