PAGE 4-91118824, VER. 00
V
te
If AVAPS is enabled, the target tidal volume can be set from 200-1500 ml in 10 ml
increments. The default setting is 200. This setting is available only on the BiPAP AVAPS
device.
IPAP Max If AVAPS is enabled, the IPAP Max can be set from 4-25 in 1.0 increments. The default
setting is 12. This setting is available only on the BiPAP AVAPS device.
IPAP Min If AVAPS is enabled, IPAP Min setting from 4-25 in 1.0 increments. The default setting is
12. This setting is available only on the BiPAP AVAPS device.
IPAP When AVAPS is not enabled, IPAP can be set to 4-25 in 1.0 increments. The default
setting is 12.
EPAP In any mode except CPAP, EPAP can be set from 4-IPAP in 1.0 increments. The default
setting is 4.
BPM In S/T or PC mode, Breaths Per Minute setting is from 0-30 in 1.0 BPM increments. In T
mode, the minimum BPM setting is 4. The default is 10.
T
i
In S/T, T, or PC mode, the Inspiratory Time can be set from 0.5 to 3 seconds in 0.1
increments. The default setting is 1.0.
Rise Time
Control
If set to “Yes,” users can not adjust the Rise Time. If set to “no,” users can adjust the Rise
Time.
Rise Time Rise time is the time it takes for the device to change from EPAP to IPAP. The default
setting is 1.
Ramp Time Ramp Time can be set from 0 (off) - 45 minutes in 5-minute increments. The default setting
is 0.
Ramp Start
Pressure
Ramp Start Pressure can be set from 4 to the CPAP or EPAP setting (depending on your
therapy mode) in increments of 1. The default is 4.
Apnea The Apnea alarm detects the cessation of spontaneous breathing. Enable or disable the
Apnea alarm by choosing 0 (off), 10, 20, or 30 seconds. The alarm will sound when the
time between patient-triggered breaths is greater or equal to the specified apnea alarm
setting. The default is 0.
Patient
Disconnect
Enable or disable the Patient Disconnect alarm by choosing 0, 15, or 60 seconds. The
alarm will sound when a large, continuous air leak is detected in the circuit for more than
the specified alarm setting. The default is 0.
MinVent Enable or disable the Low Minute Ventilation alarm by choosing 0 (off) to 99 lpm in 1.0
increments. The alarm will sound when the calculated minute ventilation is less than or
equal to the specified setting. The default is 0 (off).
Low Vte When AVAPS is enabled, enable or disable the Low Tidal Volume alarm by choosing 0
(off) or 1 (on). If enabled, the alarm will sound when the measured patient pressure has
reached the IPAP max setting and the estimated exhaled tidal volume is less than or equal
to 90% of the target tidal volume. The default is 0 (off). BiPAP AVAPS device only.
Backlight Enable or disable the button backlights with this setting.
Language Select the language to display on-screen, or select “Icon” to display icons on-screen.
hPa/cmH
2
O Select the units of pressure that are displayed on-screen, “hPa” or “cmH
2
O”.