PAGE 8-111118824, VER. 00
FIGURE 8-11: START MENU
4. Follow the on-screen prompts to continue with the test. When prompted, connect the Flow Control
Valve and Flow Meter to the device, as shown in Figure 8-12.
F
IGURE 8-12: FLOW METER AND FLOW VALVE
NOTE
At the beginning of the test, a prompt to conduct a Humidifier test
appears. Respironics suggests conducting the Humidifier test if the
device is expected to be used with a Humidifier.
To PC
Flow Meter
Valve