PAGE 7-4
1118824, VER. 00
7.2.2 REPLACING THE TANK TOP SEAL
FIGURE 7-2: TANK TOP SEAL
TO REMOVE THE TANK TOP SEAL:
1. Remove the Humidifier Tank Assembly. Refer to Replacing the Water Chamber Assembly.
2. Remove the Tank Top Seal from the Flip Lid Assembly.
TO INSTALL THE TANK TOP SEAL:
1. Lift the Flip Lid Assembly.
2. Press the Tank Top Seal onto the Patient Outlet Swivel Clip.
Included in Kit Tools Required
Tank Top Seal
None