PAGE 5-31118824, VER. 00
FIGURE 5-3: START MENU
5. Verify that a “Error Log Cleared” confirmation window appears.
5.3 ERROR CODES
The four Error Types are described as follows:
ERROR TYPE DESCRIPTION
Vent Inop The error information is recorded in NVRAM and the unit is placed into Safe State. The
only functionality available to the user is serial communication, turning off the audible
alarm via a key press and removing power.
Reboot The error information is recorded in NVRAM and the unit is Rebooted.
High Priority These alarms require immediate operator response. The alarm signal consists of a red
LED and a high priority sound. The display has the message ALARM at the top of the
screen.
Medium Priority These alarms require prompt operator response. The alarm signal consists of a yellow
LED and a medium priority sound. The display has the message ALARM at the top of
the screen.
Low Priority These alarms require operator awareness. The alarm signal consists of a yellow LED
and a low priority sound. The display has the message ALARM at the top of the screen.
Log Only The error information is recorded in NVRAM and the unit continues to operate without
noticeable alteration.