The following tables list the error codes for the Philips Respironics NIV devices.
CODE DESCRIPTION TYPE ACTION
E-0 No Error - you should only see this code in the
error log when the log entry is empty - never in
the middle of an otherwise full log.
N/A None
E-1 Software entered state that causes Device In-
operative
Stop Replace the Therapy PCA
E-2 Failed the destructive power on self test of in-
ternal RAM
Reboot Replace the Therapy PCA
E-3 Reserved for future use. Log Only None
E-4 Reserved for future use. Log Only None
E-5 Reserved for future use. Log Only None
E-6 Software Failure Log Only Replace the Therapy PCA
E-7 Failed the power on self test of the watchdog
timer
Reboot Replace the Therapy PCA
E-8 Failed the power on self test of the watchdog
timer.
Reboot Replace the Therapy PCA
E-9 One of the tasks in the background thread is
preventing the remaining tasks from executing.
Reboot Replace the Therapy PCA
E-10 During RASP operation response exceeded
time limit
Log Only Replace the Therapy PCA
E-11 During Power-On tests, an error occurred while
testing the watchdog timer.
Reboot Replace the Therapy PCA
E-12 Reserved for future use Log Only None
E-13 Reserved for future use Log Only None
E-14 Processor not able to establish communication
with the Flow Sensor
Vent Inop Replace the Therapy PCA
E-15 Reserved for future use Log Only None
E-16 Reserved for future use Log Only None
E-17 Reserved for future use Log Only None
E-18 Reserved for future use Log Only None
E-19 During spinup of the Blower, the drive detected
that the rotor was not spinning or wires were
broken.
Reboot In the following order:
Reseat Blower Con-
nectors
Replace Blower Assy
Replace Therapy
PCA
E-20 Processor is unable to communicate with
Pressure Sensor
Vent Inop Replace Therapy PCA
E-21 Processor not able to establish communication
with the Flow Sensor
Reboot Clear error log and retest. If re-
peats, replace Therapy PCA