PAGE 6-11118824, VER. 00
CHAPTER 6: NIV DEVICE REPAIR & REPLACE
This Chapter illustrates the names and locations of the replaceable components in the Philips Respironics
BiPAP AVAPS and BiPAP S/T devices. If repair or replacement procedures are performed, the device must be
run-in for a minimum of eighteen (18) minutes, and tested to verify its proper operation. Refer to Chapter 8 for
Testing Procedures.
WARNING
To prevent electrical shock, disconnect the electrical supply before
attempting to make any repairs to these devices.
CAUTION
Components used in this device are subject to damage from static electricity. Repairs made to this device
must be performed only in an antistatic, Electro-Static Discharge (ESD) protected environment.