EasyManua.ls Logo

Philips BiPAP AVAPS - 6.1.10 Replacing the Blower Assembly and;or the Blower Outlet Bellows; Figure 6-14: Blower Assembly and Blower Outlet Bellows; 1. Remove the UI Knob. Refer to Replacing the User Interface (UI) Knob.; 2. Remove the Top Cover. Refer to Replacing the Top Cover.

Philips BiPAP AVAPS
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PAGE 6-16
1118824, VER. 00
6.1.10 REPLACING THE BLOWER ASSEMBLY AND/OR THE BLOWER OUTLET BELLOWS
FIGURE 6-14: BLOWER ASSEMBLY AND BLOWER OUTLET BELLOWS
TO REMOVE THE BLOWER ASSEMBLY/OUTLET BELLOWS:
1. Remove the UI Knob. Refer to Replacing the User Interface (UI) Knob.
2. Remove the Top Cover. Refer to Replacing the Top Cover.
3. Remove the Main PCA. Refer to Replacing the Main PCA.
4. Remove the Blower Cap.Refer to Replacing the Blower Cap
5. Lift the Blower Assembly out of the Blower Housing. Remove the Blower Outlet Bellows from the
Blower Assembly.
To install the Blower Assembly/Outlet Bellows:
1. Place the Outlet Bellows onto the Blower Assembly.
Included in Kit Tools Required
Blower Assembly
Blower Outlet Bellows
T8 Torx screwdriver
T15 Torx screwdriver
Blower and Blower
Outlet Bellows
Removed from Device

Table of Contents

Other manuals for Philips BiPAP AVAPS

Related product manuals