Philips BiPAP AVAPS - 4. Remove the Inlet Seal from the Dry Box Assembly.; Figure 7-7: Dry Box Seal Removal; 1. Install The Inlet Seal onto the Dry Box Assembly if necessary.; 2. Slide the Dry Box Assembly with Inlet Seal into its mounting location in the Humidifier Bottom Housing. Verify that the Dry Box Assembly with Inlet Seal are secured and do not fall out of the Housing.

Philips BiPAP AVAPS
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PAGE 7-8
1118824, VER. 00
4. Remove the Inlet Seal from the Dry Box Assembly.
FIGURE 7-7: DRY BOX SEAL REMOVAL
TO INSTALL THE DRY BOX ASSEMBLY/HUMIDIFIER INLET SEAL:
1. Install The Inlet Seal onto the Dry Box Assembly if necessary.
2. Slide the Dry Box Assembly with Inlet Seal into its mounting location in the Humidifier Bottom
Housing. Verify that the Dry Box Assembly with Inlet Seal are secured and do not fall out of the
Housing.
Do not press firmly on the
screwdriver handle. Very light
pressure is needed to release the
Dry Box Assembly.