6.1.14 REPLACING THE BLOWER HOUSING
FIGURE 6-19: BLOWER HOUSING
To remove the Blower Housing:
1. Remove the UI Knob. Refer to Replacing the User Interface (UI) Knob.
2. Remove the Top Cover. Refer to Replacing the Top Cover.
3. Remove the Main PCA. Refer to Replacing the Main PCA.
4. Remove the Blower Cap. Refer to Replacing the Blower Cap.
5. Remove the Blower Assembly. Refer to Replacing the Blower Assembly and/or the Blower Outlet
Bellows.
6. Remove the Right Panel Assembly. Refer to Replacing the Right Panel Assembly.
7. Remove the Humidifier Cable. Refer to Replacing the Humidifier Cable.
8. Release the four latches that secure the Blower Housing inside the Bottom Enclosure.
9. Lift the Blower Housing out of the Bottom Enclosure.
To install the Blower Housing:
1. Place the Humidifier Cable into the Bottom Enclosure. Be sure the Humidifier Cable is properly
seated in its mounting location.
2. Press the Blower Housing into its mounting location in the Bottom Enclosure. Be sure that all four
locking tabs secure the Housing.
3. Assemble the remainder of the device as instructed in previous sections.
Included in Kit Tools Required
Blower Housing
T8 Torx screwdriver
T15 Torx screwdriver