3.3 ELECTROMAGNETIC IMMUNITY
This device is intended for use in the electromagnetic environment specified below. Use, service, and testing of
the device should be performed in such an environment.
GUIDANCE & MANUFACTURER’S DECLARATION - ELECTROMAGNETIC IMMUNITY
IMMUNITY TEST
IEC 60601 TEST
LEVEL
COMPLIANCE
EMC ENVIRONMENT
GUIDANCE
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast
Transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for I/O lines
±2 kV for supply mains
±1 kV for I/O lines
Mains power quality should
be that of a typical home or
hospital environment.
Surge
IEC 61000-4-5
±1 kV Differential Mode
±2 kV Common Mode
±1 kV differential mode
±2 kV common mode
Mains power quality should
be that of a typical home or
hospital environment.
Voltage dips, short
interruptions, and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
<5% U
T
(>95% dip in U
T
) for 0.5
cycle
40% U
T
(60% dip in U
T
) for 5
cycles
70% U
T
(30% dip in U
T
) for 25
cycles
<5% U
T
(>95% dip in U
T
) for 5
sec
Mains power quality should
be that of a typical home or
hospital environment. If the
user of the device requires
continued operation during
power mains interruptions, it
is recommended that the
device be powered from an
uninterruptible power supply
or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical hospital
or home environment.
NOTE: U
T
is the AC mains voltage prior to application of the test level.