EasyManua.ls Logo

Philips BiPAP AVAPS - 6.1.18 Replacing the Bottom Enclosure; Figure 6-23: Bottom Enclosure; 1. Remove the UI Knob. Refer to Replacing the User Interface (UI) Knob.; 2. Remove the Top Cover. Refer to Replacing the Top Cover.

Philips BiPAP AVAPS
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PAGE 6-26
1118824, VER. 00
6.1.18 REPLACING THE BOTTOM ENCLOSURE
FIGURE 6-23: BOTTOM ENCLOSURE
TO REPLACE THE BOTTOM ENCLOSURE:
1. Remove the UI Knob. Refer to Replacing the User Interface (UI) Knob.
2. Remove the Top Cover. Refer to Replacing the Top Cover.
3. Remove the Main PCA. Refer to Replacing the Main PCA.
4. Remove the Blower Cap. Refer to Replacing the Blower Cap.
5. Remove the Blower Assembly. Refer to Replacing the Blower Assembly and/or the Blower Outlet
Bellows.
6. Remove the Right Panel Assembly. Refer to Replacing the Right Panel Assembly.
7. Remove the Humidifier Cable. Refer to Replacing the Humidifier Cable.
8. Remove the Blower Housing. Refer to Replacing the Blower Housing.
9. Remove the Sound Abatement Foam. Refer to Replacing the Sound Abatement Foam.
10. Remove the Air inlet Seal. Refer to Replacing the Air Inlet Seal.
11. Install all the components into the new Bottom Enclosure and assemble the device as instructed in
previous sections.